Additional narrative: (b)(4).Date of non-union is not known.Date of implant reported as approximately one year prior to removal.Complainant part is not expected to be returned for manufacturer review/investigation.The investigation could not be completed; no conclusion could be drawn, as no product was received.Dhr review: (b)(4), manufacturing date: 01-may-2014, expiration date: 31-mar-2023.Part #: 04.003.860s, lot#: 7671625 (sterile) ¿ 15mm ti-lateral entry femoral recon nail-ex/400mm/rt - sterile.Quantity 6.Inspection sheet for inspection dimensional & final inspection met inspection acceptance criteria.Component parts reviewed: part 21069, lot: 7506606 for titanium received from (b)(4).(b)(4) certificate of test for titanium meets specification.Raw material receiving/putaway checklist meet specification.No ncrs were generated during production.Review of the device history record(s) showed that there were no issues during the manufacture of the product that would contribute to this complaint condition.¿sterility documentation was reviewed and determined to be conforming.¿ device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|