| Catalog Number |
CCS29 |
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
09/22/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.The lot history records were reviewed and the manufacturing criteria were met prior to the release of the lot.
|
| |
|
Event or Problem Description
|
|
It was reported that during the endoscopic low rectal resection, after fired, found only with good cut line, but the staples malformed.Suture was used to sew the wound.There were no adverse consequences to the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Batch # m54h3u.Device evaluation: the analysis results found that the ccs29 device arrived and with no apparent damage.The breakaway washer was present and uncut and the device was fully loaded with staples, indicating that the device had not being fired.The device was tested for functionality with the returned washer and it fired and formed all the staples as well as completely cut the test media and the breakaway washer without incident.The staple line was complete and the staples were noted to have the proper b-formed shape.The event described could not be confirmed as the device performed without any difficulties noted.The batch history record was reviewed and identified a defect/issue but was not related to the reported incident were found.In addition, no protocols or ncr related to the complaint, were found during the manufacturing process.
|
| |
|