• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEXCOM, INC. DEXCOM G5 MOBILE CONTINUOUS GLUCOSE MONITORING SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEXCOM, INC. DEXCOM G5 MOBILE CONTINUOUS GLUCOSE MONITORING SYSTEM Back to Search Results
Model Number MT22719-PNK
Medical Device Problem Code Device Operates Differently Than Expected (2913)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 12/18/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).
 
Event or Problem Description
Dexcom was made aware on (b)(4) 2017, that on (b)(6) 2017, the patient stated the receiver stops showing values if it is not calibrated immediately after being prompted to calibrated.No medical intervention was reported.Additional event or patient information is not available.The device has been received for evaluation.  a follow-up report will be submitted once the evaluation is complete. .
 
Additional Manufacturer Narrative
(b)(4).
 
Event or Problem Description
The complaint device was returned for evaluation.The device was visually inspected and no defect was found.Receiver was charged and booted up.Functional testing was performed and there was no failure detected related to the complaint.A pairing test was performed with a known good transmitter and passed.A review of the downloaded data log confirmed the customer's complaint.A root cause could not be determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DEXCOM G5 MOBILE CONTINUOUS GLUCOSE MONITORING SYSTEM
Common Device Name
CONTINUOUS GLUCOSE MONITOR
Manufacturer (Section D)
DEXCOM, INC.
6340 sequence drive
san diego CA 92121
MDR Report Key7189998
Report Number3004753838-2018-05284
Device Sequence Number111193
Product Code PQF
UDI-Device Identifier30386270000263
UDI-Public30386270000263
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
P120005
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 12/18/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberMT22719-PNK
Device Catalogue NumberSTK-GF-PNK
Device Lot Number5227077
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/27/2017
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/18/2017
Supplement Date Received by Manufacturer01/18/2018
Initial Report FDA Received Date01/15/2018
Supplement Report FDA Received Date01/30/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age7 YR
-
-