| Model Number |
MT22719-PNK |
| Medical Device Problem Code |
Device Operates Differently Than Expected (2913)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
12/18/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).
|
| |
|
Event or Problem Description
|
|
Dexcom was made aware on (b)(4) 2017, that on (b)(6) 2017, the patient stated the receiver stops showing values if it is not calibrated immediately after being prompted to calibrated.No medical intervention was reported.Additional event or patient information is not available.The device has been received for evaluation. a follow-up report will be submitted once the evaluation is complete. .
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).
|
| |
|
Event or Problem Description
|
|
The complaint device was returned for evaluation.The device was visually inspected and no defect was found.Receiver was charged and booted up.Functional testing was performed and there was no failure detected related to the complaint.A pairing test was performed with a known good transmitter and passed.A review of the downloaded data log confirmed the customer's complaint.A root cause could not be determined.
|
| |
|