| Model Number |
PASSEO-35 HP 4/40/75 |
| Medical Device Problem Code |
Inflation Problem (1310)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
12/14/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Top assembly batch history record was reviewed.The units were processed to the correct validated parameters.All units were 100% visually inspected and leak tested by trained operators to a rated pressure of 27 atm.The material and components used in this device were determined to be consistent with all materials used in the manufacture of all passeo-35 hp products.Sub assembly balloons are subject to burst and compliance testing as per batch release requirements.The sub-assembly balloon batch used met all batch release requirements.There were no non conformances, deviations, capa's or unusual anomalies associated with the impacted batch.Further investigation will be carried out once the defective unit is returned.
|
| |
|
Event or Problem Description
|
|
The physician tried to inflate the passeo-35hp balloon up to 20 atm in avf with approximately 70% stenosis.During the attempt, a possible balloon tear was observed under angiography; the balloon had failed to fully inflate and the shape of the balloon looked limp, even showing slight leak of the inflation medium.After withdrawing the defect device, a new passeo-35hp of the same size was used to complete the treatment.
|
| |
|
Additional Manufacturer Narrative
|
|
Top assembly batch history record was reviewed.The units were processed to the correct validated parameters.All units were 100% visually inspected and leak tested by trained operators to a rated pressure of 27 atm.The material and components used in this device were determined to be consistent with all materials used in the manufacture of all passeo-35 hp products.Sub assembly balloons are subject to burst and compliance testing as per batch release requirements.The sub-assembly balloon batch used met all batch release requirements.There were no non conformances, deviations, capa's or unusual anomalies associated with the impacted batch.The defective unit was returned for investigation, the part and batch details on the returned unit were confirmed as correct and as per the complaint notification.The catheter shaft unit was found to be kinked, close to the manifold, no other visual anomalies were noted on the device.The unit was inflated using a syringe.The balloon was fully inflated displaying the correct balloon profile/form.Inspection was carried out under 10x magnification and no balloon tears or leaks were detected.However, during the leak testing a leak was noted emanating from the glue port holes in the manifold of the device.This finding was not documented in the initial complaint report, it therefore cannot be confirmed if this failure mode was present during use or if it was introduced after the initial complaint was received.The additional failure mode does not increase the criticality of the complaint grade or require a re-assessment of the initial complaint grade assigned.
|
| |
|
Event or Problem Description
|
|
The physician tried to inflate the passeo-35hp balloon up to 20 atm in avf with approximately 70% stenosis.During the attempt, a possible balloon tear was observed under angiography; the balloon had failed to fully inflate and the shape of the balloon looked limp, even showing slight leak of the inflation medium.After withdrawing the defect device, a new passeo-35hp of the same size was used to complete the treatment.
|
| |
|