| Model Number |
MT22719-PNK |
| Medical Device Problem Code |
Device Displays Incorrect Message (2591)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
12/17/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).
|
| |
|
Event or Problem Description
|
|
Dexcom was made aware on (b)(4) 2017, that on (b)(6) 2017, the receiver displayed err121.No additional patient or event information is available.No product or data was returned for evaluation.The complaint confirmation of the error icon could not be determined.A root cause could not be determined.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).
|
| |
|
Event or Problem Description
|
|
The complaint receiver was returned for evaluation.The device was visually inspected and no defect was found.The receiver failed to charged and reboot with the "call tech support alarm" displayed.The receiver log was reviewed firmware errors were observed.The reported event of an error icon display was confirmed during log review.A root cause could not be determined.
|
| |
|