• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TERUMO MEDICAL CORPORATION RADIFOCUS INTRODUCER II KIT; INTRODUCER, CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TERUMO MEDICAL CORPORATION RADIFOCUS INTRODUCER II KIT; INTRODUCER, CATHETER Back to Search Results
Model Number N/A
Medical Device Problem Code Leak/Splash (1354)
Health Effect - Clinical Codes No Consequences Or Impact To Patient (2199); Blood Loss (2597)
Date of Event 12/20/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Implanted date: device was not implanted.Explanted date: device was not explanted.The actual device has been returned for evaluation.The investigation is currently ongoing.A follow up report will be submitted once the investigation is complete.For this reason, evaluation code has been referenced in the conclusions.A review of the device history record of the product code/lot# combination was conducted with no findings.
 
Event or Problem Description
The user facility reported during an evar procedure the physician noted that the valve of the sheath was leaking abnormally.Blood loss was less than 250cc.The patient was reported to be fine.The procedure was completed.
 
Additional Manufacturer Narrative
This report is being submitted as follow up no.1 to provide the completed investigation.Two unused 8fr ik devices associated with lot vf03 were returned to for product evaluation.The returned devices were subjected to visual analysis.No anomalies were detected with the packaging.All seals were intact.There are no anomalies noted with either the sheaths or the dilators.Leak testing was performed on one of the unused sealed devices and no functional anomalies were detected.Based on the returned unused samples, there is no evidence that this event was related to a device defect or malfunction.The exact cause of the reported event cannot be definitively determined based on the available information.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RADIFOCUS INTRODUCER II KIT
Common Device Name
INTRODUCER, CATHETER
Manufacturer (Section D)
TERUMO MEDICAL CORPORATION
950 elkton blvd.
elkton MD 21921
MDR Report Key7192906
Report Number1118880-2017-00099
Device Sequence Number1988554
Product Code DYB
UDI-Device Identifier00389701007663
UDI-Public00389701007663
Combination Product (Y/N)N
Initial Reporter StateNE
Initial Reporter CountryUS
PMA/510(K) Number
K954234
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 01/16/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/31/2019
Device Model NumberN/A
Device Catalogue NumberRSS804
Device Lot NumberVF03
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/10/2018
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 12/20/2017
Supplement Date Received by Manufacturer01/27/2018
Initial Report FDA Received Date01/16/2018
Supplement Report FDA Received Date01/29/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age64 YR
Patient Weight99
-
-