• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: GUIDANT CRM CLONMEL IRELAND ACCOLADE; IMPLANTABLE PULSE GENERATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

GUIDANT CRM CLONMEL IRELAND ACCOLADE; IMPLANTABLE PULSE GENERATOR Back to Search Results
Model Number U125
Device Problems High impedance (1291); Over-Sensing (1438); Ambient Noise Problem (2877)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 06/01/2017
Event Type  malfunction  
Manufacturer Narrative
As no further information concerning this report is expected, our investigation is complete.This investigation will be updated should further information be provided.
 
Event Description
Boston scientific received information that this pacemaker exhibited noise and oversensing along with increasing pacing impedance.Recently, the lead safety switch (lss) was triggered due to high out of range impedances, measuring greater than 2,000 ohms.No adverse patient effects were reported.The device was programmed to pace only in the the ventricle due to the noise.The patient will continue to be monitored.This product remains in-service.The patient's right ventricular lead is a non-boston scientific product.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACCOLADE
Type of Device
IMPLANTABLE PULSE GENERATOR
Manufacturer (Section D)
GUIDANT CRM CLONMEL IRELAND
guidant ireland
clonmel, tipperary ireland
Manufacturer (Section G)
GUIDANT CRM CLONMEL IRELAND
guidant ireland
clonmel, tipperary ireland
Manufacturer Contact
tim degroot
4100 hamline ave. n
st. paul, MN 
6515826168
MDR Report Key7192993
MDR Text Key97672826
Report Number2124215-2017-20838
Device Sequence Number1
Product Code NKE
UDI-Device Identifier00802526559389
UDI-Public(01)00802526559389(17)20170909
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030005/S113
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Medical Equipment Company Technician/Representative
Type of Report Initial
Report Date 11/02/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/16/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date09/09/2017
Device Model NumberU125
Other Device ID NumberVALITUDE
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? Yes
Date Manufacturer Received11/02/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/09/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1