MAUDE Adverse Event Report: TORNIER S.A.S. AEQUALIS PERFORM GLENOID; SHOULDER JOINT POLYMER PROSTHESIS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TORNIER S.A.S. AEQUALIS PERFORM GLENOID; SHOULDER JOINT POLYMER PROSTHESIS Back to Search Results
Model Number SIZE L60
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 12/15/2017
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
This is the initial report submitted regarding this surgical event and medical device.
 
Event or Problem Description
The surgeon prepared the glenoid for an l60 cortiloc glenoid and the trial seated flush with the bone.When he went to seat the final implant, it would not sit flush with the bone.He tried to get it seated without success and ultimately decided to pull it out and it came out fairly easily.Upon inspection, the surgeon observed the inferior-posterior peg was bent and concluded this was the issue.The surgeon removed the cement and re-prepped the glenoid for an l50 which added 20 minutes approximately to the surgery.
 
Additional Manufacturer Narrative
This is the final report submitted regarding this surgical event and medical device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AEQUALIS PERFORM GLENOID
Common Device Name
SHOULDER JOINT POLYMER PROSTHESIS
Manufacturer (Section D)
TORNIER S.A.S.
161 rue lavoisier
montbonnot saint-martin, 38330
FR  38330
MDR Report Key7193172
Report Number3000931034-2018-00001
Device Sequence Number19298712
Product Code KWS
UDI-Device Identifier03700386937538
UDI-Public03700386937538
Combination Product (Y/N)N
Initial Reporter StateIN
Initial Reporter CountryUS
PMA/510(K) Number
K111902
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Reporter Type Manufacturer
Report Source other
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 12/18/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Physician
Device Expiration Date02/15/2022
Device Model NumberSIZE L60
Device Catalogue NumberDWE423
Device Lot NumberAA9112
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 12/18/2017
Supplement Date Received by Manufacturer12/18/2017
Initial Report FDA Received Date01/16/2018
Supplement Report FDA Received Date07/23/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age63 YR
-
-