Model Number 4543
Medical Device Problem Codes Failure to Capture (1081); Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 11/09/2017
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
At this time, the lead has not been returned.If the product is returned, analysis will be performed and this report would be updated at that time.
 
Event or Problem Description
Boston scientific received information that this left ventricular (lv) lead was explanted one day after it was implanted due to dislodgement.At post check there was no capture at max output.The lead was not replaced.No additional adverse patient effects were reported.
 
Additional Manufacturer Narrative
This lead has been returned.Boston scientific has concluded it is unlikely that lead characteristics caused or contributed to the reported clinical observations.As no further information concerning this report is expected, our investigation is complete.This investigation will be updated should further information be provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EASYTRAK 2
Common Device Name
IMPLANTABLE LEAD
Manufacturer (Section D)
CPI - DEL CARIBE
guidant puerto rico b. v.
dorado PR
MDR Report Key7193729
Report Number2124215-2017-21900
Device Sequence Number19329893
Product Code NIK
UDI-Device Identifier00802526579813
UDI-Public00802526579813
Combination Product (Y/N)N
Initial Reporter StateMN
Initial Reporter CountryUS
PMA/510(K) Number
P010012/S024
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2017
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/13/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date05/15/2019
Device Model Number4543
Other Device ID NumberEASYTRAK 2
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/04/2017
Was the Report Sent to FDA? Yes
Initial Date Received by Manufacturer 11/10/2017
Supplement Date Received by Manufacturer12/13/2017
Initial Report FDA Received Date01/16/2018
Supplement Report FDA Received Date01/19/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
4543; 7740; L331
Outcome Attributed to Adverse Event Hospitalization; Life Threatening; Required Intervention;
Patient Age82 YR
-
-