COOK INC ADVANCE 14 LP LOW PROFILE BALLOON CATHETER; DQY CATHETER, PERCUTANEOUS Back to Search Results
Catalog Number PTAX4-14-170-2.5-20
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 01/05/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).This report includes information known at this time. a follow-up report will be submitted should additional relevant information become available.
 
Event or Problem Description
It was reported that during treatment of the patient's superficial femoral artery (sfa) lesion, the advance 14 lp low profile balloon catheter was advanced to the patient's lesion and ruptured when it was attempted be inflated at 8 atmosphere (atm).Access to the lesion was gained from brachial artery.Another device was used to successfully complete the procedure.A section of the device did not remain inside the patient¿s body.According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.
 
Additional Manufacturer Narrative
Investigation - evaluation: a review of the complaint history, drawings, device history record, instructions for use (ifu), manufacturing instructions, specifications, and quality control data was conducted during the investigation.The complaint device was not returned; therefore, no physical examination could be performed.However, a document-based investigation was performed.There is no evidence to suggest the finished product was not made to specifications.Numerous design verification and validation activities have been performed to ensure that this device meets design requirement, each balloon is 100% inspected and verified prior to release.A review of the device history record of the finished product shows no nonconforming events that would contribute to this failure mode.It should also be noted, there were no other reported complaints for this lot number.Based on the information provided, no product returned, and the results of our investigation; a definitive root cause could not be determined.Per the quality engineering risk assessment, no further action is required.Appropriate personnel have been notified and monitoring will continue to be performed for similar complaints.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADVANCE 14 LP LOW PROFILE BALLOON CATHETER
Common Device Name
DQY CATHETER, PERCUTANEOUS
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
MDR Report Key7194391
Report Number1820334-2018-00060
Device Sequence Number1863744
Product Code DQY
UDI-Device Identifier10827002503264
UDI-Public(01)10827002503264(17)190208(10)6718864
Combination Product (Y/N)N
PMA/510(K) Number
K090822
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional,u
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/23/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberPTAX4-14-170-2.5-20
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 01/05/2018
Supplement Date Received by Manufacturer01/22/2018
Initial Report FDA Received Date01/16/2018
Supplement Report FDA Received Date01/23/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-