The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.The following were reviewed as part of this investigation: patient severity, frequency analysis, applicable previous investigation(s), sample (if available), applicable fmea documents, labeling, and applicable manufacture records.Based on a review of this information, the following was concluded: the complaint of a frayed guidewire is inconclusive due to poor sample condition.One photo sample of a microez introducer package label was returned for evaluation.An initial visual observation of the photo showed the product labeling of a microez introducer kit with ref#: 0738945 and lot#: rebu0288.The device is not visible in the photo provided.Based on the description of the reported complaint, possible contributing factors include kinked guidewire and needle damage to guidewire.As the complaint of a frayed guidewire cannot be confirmed based on the photo sample provided for evaluation, the complaint is inconclusive, poor sample condition.The ifu states ¿caution: if the guidewire must be withdrawn while the needle is inserted, remove both the needle and the wire as a unit to prevent the needle from damaging or shearing the guidewire.¿ a lot history review (lhr) of rebu0288 showed no other similar product complaint(s) from this lot number.
|