| Model Number |
S403 |
| Medical Device Problem Code |
Device Operates Differently Than Expected (2913)
|
| Health Effect - Clinical Code |
Syncope (1610)
|
| Date of Event |
11/24/2017
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The local area sales representative was contacted for additional information.At this time, no further information is available.Should additional information become available this report will be updated.
|
| |
|
Event or Problem Description
|
|
Boston scientific received information that the patient implanted with this pacemaker experienced a syncopal episode.Review of stored device memory noted several atrial tachy response (atr) episodes that was suspected to be a potential new condition.It was unknown if these episodes correlated with the syncope.No additional adverse patient effects were reported.This product remains implanted and in service.
|
| |
|
Additional Manufacturer Narrative
|
|
Upon receipt at our post market quality assurance laboratory, a thorough evaluation of the device was performed.A visual inspection of the device header and case noted no anomalies.The device was then exposed to simulated heart load conditions, and the pacing, and sensing functions were tested.Impedance testing was completed and all measurements were within normal limits.The device operated appropriately with no interruptions in therapy output at the returned programmed settings.A series of electrical tests was also performed, and again, normal device function was observed.Analysis did not identify any device characteristics that would have caused or contributed to the reported clinical observations.
|
| |
|
Event or Problem Description
|
|
Additional information was received that the pacemaker was explanted and replaced for reported normal battery depletion approximately two months later.
|
| |
|