ANIMAS CORPORATION ANIMAS VIBE; INSULIN INFUSION PUMP Back to Search Results
Medical Device Problem Codes Partial Blockage (1065); Device Displays Incorrect Message (2591)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The pump has been returned to animas.Evaluation has not yet been completed.When evaluation is complete a supplemental report will be filed.No conclusion can be made at this time.
 
Event or Problem Description
On (b)(6) 2017, the reporter contacted animas, alleging a occlusion (frequent/persistent) issue.There was no indication that the device caused or contributed to an adverse event.This complaint is being reported because the issue may result in a long-term cessation of insulin delivery if the user is unable to resolve the alarm.
 
Additional Manufacturer Narrative
Follow-up #1: date of submission 26-jan-2018 - device evaluation: the device has been returned and evaluated by product analysis on 22-jan-2018 with the following findings: during the investigation, multiple occlusion alarms were observed in the black box.The force at the time of the occlusions was found to be high.The rewind, load, and prime steps were performed correctly with no alarms.Force calibration testing passed within specifications.The pump was exercised for 24 hours with no occlusions occurring.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ANIMAS VIBE
Common Device Name
INSULIN INFUSION PUMP
Manufacturer (Section D)
ANIMAS CORPORATION
200 lawrence dr
west chester PA 19380 3428
MDR Report Key7196042
Report Number2531779-2018-01075
Device Sequence Number1496861
Product Code OYC
UDI-Device Identifier10840406100266
UDI-Public10840406100266
Combination Product (Y/N)N
Initial Reporter StateNV
Initial Reporter CountryUS
PMA/510(K) Number
P130007
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Other Caregivers
Type of Report Initial,Followup
Report Date (Section B) 12/22/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Other Device ID Number1-MCBC-3762
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/15/2018
Was the Report Sent to FDA? Yes
Device Age9 MO
Initial Date Received by Manufacturer 12/22/2017
Supplement Date Received by Manufacturer12/22/2017
Initial Report FDA Received Date01/16/2018
Supplement Report FDA Received Date01/26/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age32 YR
Patient Weight193
-
-