• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OBERDORF SYNTHES PRODUKTIONS GMBH TI CLICK'X® LOCKING CAP FOR TI 3-D HEAD; ORTHOSIS,SPONDYLOSTHESIS SPINAL FIXATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OBERDORF SYNTHES PRODUKTIONS GMBH TI CLICK'X® LOCKING CAP FOR TI 3-D HEAD; ORTHOSIS,SPONDYLOSTHESIS SPINAL FIXATION Back to Search Results
Model Number 498.570
Medical Device Problem Code Fitting Problem (2183)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 12/18/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Patient¿s identifier and weight are unknown.Device remains implanted in patient; as such explant date is not applicable.Complainant device is not expected to be returned for manufacturer review/investigation.The investigation could not be completed; no conclusion could be drawn, as no product was received.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Device report from synthes europe reports an event in (b)(6) as follows: it was reported that during a surgery performed on (b)(6) 2017, the surgeon noted that the screwdriver available for use would not work with the click´x system + t-pal implants being placed in a patient.Reportedly, the patient was anesthetized and as the surgeon proceeded to place the six click´x screws he had to force the screwdriver so that the first screw was placed correctly and locked with the corresponding click x locking cap.The screwdriver did not work at all and the surgeon was unable to correctly secure the other five click x locking caps.Additionally reported, the surgeon received with the set, two straight rods and not two curved rods as expected.As such, he had to manually adapt the bar to fit the construct.The surgeon completed the surgery but only one locking cap was fit as expected, the other five did not fit (nor did the devices make the click sound as expected), the surgeon tried to fit as much as screwdriver allowed.The surgery was delayed by 45 minutes due to the intra-operative events.The patient was reportedly discharged from the clinic and is under periodic supervision.There have been no changes to the patient status and the implanted devices show now device malfunction or any adverse experience such as pain.Concomitant devices reported: 7.0mm ti click'x® pedicle screw preassembled 35mm thrd length (part # 498.994, lot number unknown, quantity 2), 6.2mm ti click'x® pedicle scr preassembled 45mm thrd length (part # 498.989, lot # unknown, quantity 4), ti click'x® locking cap for ti 3-d head (part # 498.570, lot # l227879, quantity 1).This report is for one (1) ti click'x® locking cap for ti 3-d head.This is report 5 of 5 for complaint (b)(4).
 
Additional Manufacturer Narrative
Additional narrative: patient age or date of birth not available for reporting.Dhr review was completed.Part no.: 498.570, lot no.: l227879, manufacturing location: (b)(4), release to warehouse date: 30.Nov.2016.No ncrs were generated during production.Review of the device history record(s) showed that there were no issues during the manufacture of the product that would contribute to this complaint condition.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TI CLICK'X® LOCKING CAP FOR TI 3-D HEAD
Common Device Name
ORTHOSIS,SPONDYLOSTHESIS SPINAL FIXATION
Manufacturer (Section D)
OBERDORF SYNTHES PRODUKTIONS GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
MDR Report Key7196400
Report Number8030965-2018-50407
Device Sequence Number16632220
Product Code MNH
UDI-Device Identifier10705034791081
UDI-Public(01)10705034791081(10)L227879
Combination Product (Y/N)N
Initial Reporter CountryPE
PMA/510(K) Number
K082572
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Health Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 12/18/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number498.570
Device Catalogue Number498.570
Device Lot NumberL227879
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 12/18/2017
Supplement Date Received by Manufacturer01/25/2018
02/28/2018
Initial Report FDA Received Date01/16/2018
Supplement Report FDA Received Date02/12/2018
03/06/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Weight80
-
-