AV-TEMECULA-CT NC TREK CORONARY DILATATION CATHETER Back to Search Results
Catalog Number 1012452-15
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 01/06/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).During processing of this complaint, attempts were made to obtain complete event, patient and device information.The device was not returned for analysis.The complaint investigation determined the reported difficulty was related to manufacturing issues associated with the protective sheath.On march 14, 2017, abbott vascular decided to initiate a voluntary field action for some sizes and lots of the nc trek family of dilatation catheters for difficulty to remove sheath which may lead to inflation or deflation issues.Abbott vascular performed a comprehensive investigation which included device analysis, manufacturing evaluation and trend analysis.The root cause identification was complicated by the fact that users were describing multiple symptoms when reporting the complaints.To date, the frequency of worldwide reported events for difficulties removing the protective balloon sheath, inflation and deflation has reached an actionable limit, thus abbott vascular communicated the voluntary field action to the fda on march 17, 2017 [medwatch#2024868-2017-02310].Corrective action has been implemented per site operating procedures.This product will continue to be trended.The abbott internal recall number is 2024168-3/14/2017-002r.
 
Event or Problem Description
It was reported that during a coronary artery procedure, the 3.75 x 15 mm nc trek balloon dilatation catheter (bdc) was inflated at high pressure, but failed to deflate.Attempts were made to deflate the balloon, but it would not deflate, so the device was slowly removed and the proximal shaft kinked.There was no reported adverse patient sequela or a clinically significant delay in procedure.No additional information was provided.
 
Additional Manufacturer Narrative
(b)(4).Correction: the date received by manufacturer date was filed incorrectly on the initial medwatch report as 12/18/2017, but should have been 11/12/2017.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NC TREK CORONARY DILATATION CATHETER
Common Device Name
CORONARY DILATATION CATHETER
Manufacturer (Section D)
AV-TEMECULA-CT
abbott vascular
26531 ynez road
temecula CA 92591 4628
MDR Report Key7197846
Report Number2024168-2018-00422
Device Sequence Number722570
Product Code LOX
UDI-Device Identifier08717648152030
UDI-Public08717648152030
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K110134
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Health Professional
Remedial Action Recall
Type of Report Initial,Followup
Report Date (Section B) 02/27/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/30/2019
Device Catalogue Number1012452-15
Device Lot Number60718G1
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 12/18/2017
Supplement Date Received by Manufacturer02/15/2018
Initial Report FDA Received Date01/16/2018
Supplement Report FDA Received Date02/27/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
Z-2037-2039-2017
Patient Sequence Number1
Patient Age63 YR
-
-