| Model Number |
NBD8 |
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
12/29/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.The manufacturer has received the sample and will evaluate.Results are expected soon.A lot history review (lhr) review is not possible, as no manufacturing lot number has been provided by the complainant.
|
| |
|
Event or Problem Description
|
|
It was reported that during ptc procedure the plastic dilator broke from the tip when attempting to remove it.No other information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
The following were reviewed as part of this investigation: patient severity, frequency analysis, applicable previous investigation(s), sample (if available), applicable fmea documents, labeling, and applicable manufacture records.Based on a review of this information, the following was concluded: the complaint of a damaged stylet was confirmed.One photograph was provided for investigation.The photo showed what appeared to be the connector on a navarre drainage catheter.The inside of the white connector was visible and what appeared to be a fractured end of the flexible stylet was observed within the hub.The fractured ends of the stylet were uneven and jagged.A lot history review (lhr) review is not possible, as no manufacturing lot number has been provided by the complainant.
|
| |
|
Event or Problem Description
|
|
It was reported that during ptc procedure the plastic dilator broke from the tip when attempting to remove it.No other information was provided.
|
| |
|