MAUDE Adverse Event Report: MERGE HEALTHCARE MERGE UNITY PACS; PICTURE AND ARCHIVING COMMUNIATIONS SYSTEM,

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MERGE HEALTHCARE MERGE UNITY PACS; PICTURE AND ARCHIVING COMMUNIATIONS SYSTEM, Back to Search Results
Model Number UNITY PACS R11.1.2.5
Device Problems Computer Software Problem (1112); Loss of Data (2903)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 12/19/2017
Event Type  malfunction  
Manufacturer Narrative
Merge unity pacs is designed to display a message to the user when an exam report fails to upload to the image server.In this instance, there was no evidence that the physician received the message warning that the report did not upload.It was not possible to determine if the message displayed on the screen for the physician when the report was first read and saved.Unity investigation showed the database was updated but the report did not upload to the server.Audit trail showed the exam was read, approved and saved.The exam was re-read by the physician.Station crash caused by modal conflict with (b)(6) which prevented original report upload.
 
Event Description
Merge unity pacs is a medical image and information management system that is used for viewing, selection, processing, printing, telecommunications, and media interchange of medical images from a variety of diagnostic imaging systems.On (b)(6) 2017, the pacs admin reported an exam report was stuck in apd* (approved/error) status.Users were unable to open the report.Merge healthcare unity technical support performed and investigation and it showed the customer's workstation had crashed during a report save.This workstation crash was caused by a modal conflict with (b)(6).The exam was re-read by the physician.While images are available, a finalized report not being available for subsequent review by the patient's physician could potentially result in a delay in care that could lead to harm.However, the customer did not allege any harm, serious injury or death as a result of this issue.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MERGE UNITY PACS
Type of Device
PICTURE AND ARCHIVING COMMUNIATIONS SYSTEM,
Manufacturer (Section D)
MERGE HEALTHCARE
900 walnut ridge drive
hartland WI 53029
Manufacturer (Section G)
MERGE HEALTHCARE
900 walnut ridge drive
hartland WI 53029
Manufacturer Contact
meg mucha
900 walnut ridge drive
hartland, WI 53029
MDR Report Key7198757
MDR Text Key97958839
Report Number2183926-2018-00005
Device Sequence Number1
Product Code LLZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K041935
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Other
Type of Report Initial
Report Date 12/19/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/17/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberUNITY PACS R11.1.2.5
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received12/19/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/24/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-