• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CPI - DEL CARIBE EASYTRAK 2; IMPLANTABLE LEAD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CPI - DEL CARIBE EASYTRAK 2; IMPLANTABLE LEAD Back to Search Results
Model Number 4543
Medical Device Problem Codes Capturing Problem (2891); Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 10/10/2017
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The return of the product was requested.If the product is returned, analysis will be performed and this event would be updated upon analysis completion.
 
Event or Problem Description
Boston scientific received information that this left ventricular (lv) lead exhibited high pacing thresholds.Additional information was received indicating that the cause of high pacing thresholds was due to lead was pulled back and was confirmed through fluoroscopy.In addition, the large posterior lateral vessel was not a stable position for the lv lead.This lv lead was explanted.No additional adverse patient effects were reported.
 
Additional Manufacturer Narrative
This lead has been returned.Boston scientific has concluded it is unlikely that lead characteristics caused or contributed to the reported clinical observations.As no further information concerning this report is expected, our investigation is complete.This investigation will be updated should further information be provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EASYTRAK 2
Common Device Name
IMPLANTABLE LEAD
Manufacturer (Section D)
CPI - DEL CARIBE
guidant puerto rico b. v.
dorado PR
MDR Report Key7199376
Report Number2124215-2017-19775
Device Sequence Number19046085
Product Code NIK
UDI-Device Identifier00802526579813
UDI-Public00802526579813
Combination Product (Y/N)N
Initial Reporter StateNJ
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2017
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/13/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date03/08/2018
Device Model Number4543
Other Device ID NumberEASYTRAK 2
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/31/2017
Was the Report Sent to FDA? Yes
Initial Date Received by Manufacturer 10/10/2017
Supplement Date Received by Manufacturer11/13/2017
Initial Report FDA Received Date01/17/2018
Supplement Report FDA Received Date01/25/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
4543; 7741; 7742; U125
Outcome Attributed to Adverse Event Hospitalization; Life Threatening; Required Intervention;
Patient Age62 YR
-
-