| Model Number |
4543 |
| Medical Device Problem Codes |
Capturing Problem (2891); Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
10/10/2017
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The return of the product was requested.If the product is returned, analysis will be performed and this event would be updated upon analysis completion.
|
| |
|
Event or Problem Description
|
|
Boston scientific received information that this left ventricular (lv) lead exhibited high pacing thresholds.Additional information was received indicating that the cause of high pacing thresholds was due to lead was pulled back and was confirmed through fluoroscopy.In addition, the large posterior lateral vessel was not a stable position for the lv lead.This lv lead was explanted.No additional adverse patient effects were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
This lead has been returned.Boston scientific has concluded it is unlikely that lead characteristics caused or contributed to the reported clinical observations.As no further information concerning this report is expected, our investigation is complete.This investigation will be updated should further information be provided.
|
| |
|