STRYKER SPINE-US MANTIS REDUX ROUND HANDLE NON RATCHET; ORTHOPEDIC MANUAL SURGICAL INSTRUMENT Back to Search Results
Catalog Number 48289305
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 12/20/2017
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that; or manager at the clinic, reported the following event to , sales rep : "arthrodesis procedure, patient operated by doctor.Installation and positioning of the patient was performed.While preparing the pedicle sight with jam shidi trocar, he wanted to implant the screw through the pedicle with the screwdriver, the handle broke at the rack.He had to use another handle to complete the procedure.There was a delay of about 5-10 minutes.".
 
Additional Manufacturer Narrative
Method: visual inspection, material analysis, device history review, complaint history review, risk assessment; result: the customer reported event was confirmed via visual inspection.The inner collar was found to be fractured.The fractured pieces were not returned.Manufacturing history was reviewed and no issues were identified.Upon material analysis of the returned device, it was concluded that the ratchet handle was fractured in a mixed mode of cleavage and ductile due to an applied torsional force while contacting an implant.The measured hardness and elemental composition met the requirements specified on the print.No material or manufacturing defects were found.Conclusion: the root cause of the reported event could not be determined conclusively but potential root causes include: - excessive torsional force applied by user - excess force applied to screwdriver (cantilever) - screw not properly aligned with screwdriver.
 
Event or Problem Description
It was reported that; or manager at the clinic, reported the following event to , sales rep : "arthrodesis procedure, patient operated by doctor.Installation and positioning of the patient was performed.While preparing the pedicle sight with jam shidi trocar, he wanted to implant the screw through the pedicle with the screwdriver, the handle broke at the rack.He had to use another handle to complete the procedure.There was a delay of about 5-10 minutes.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MANTIS REDUX ROUND HANDLE NON RATCHET
Common Device Name
ORTHOPEDIC MANUAL SURGICAL INSTRUMENT
Manufacturer (Section D)
STRYKER SPINE-US
2 pearl court
allendale NJ 07401
MDR Report Key7199459
Report Number0009617544-2018-00013
Device Sequence Number1044316
Product Code LXH
UDI-Device Identifier07613327266948
UDI-Public(01)07613327266948
Combination Product (Y/N)N
Initial Reporter CountryFR
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Nursing Assistant
Type of Report Initial,Followup
Report Date (Section B) 06/04/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device No Information
Device Catalogue Number48289305
Device Lot Number173904
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/13/2018
Initial Date Received by Manufacturer 12/21/2017
Supplement Date Received by Manufacturer05/07/2018
Initial Report FDA Received Date01/17/2018
Supplement Report FDA Received Date06/04/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-