| Model Number |
03821545 |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Codes |
Foreign Body In Patient (2687); No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
12/02/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that; patient was put in prone position.This was a regular operation sterilization spread towels, full clear surgical field exposure, conventional cone, open open device agent take nail, needle, the establishment of fluoroscopy location is good, choose screw implant, poor position after perspective.Surgeon removed the screw, found the screw detached and never recovered.
|
| |
|
Additional Manufacturer Narrative
|
|
No relevant manufacturing issues were identified, as all units met specifications.Excessive force was most likely applied to the implant, causing the tulip to disengage, causing the deformation seen on the implant.Possible cross threading of the screwdriver, difficult angle of the tulip head or inadequate preparation of the pathway are root causes for the user to apply excess force to remove the implant and cause the tulip to disengage.
|
| |
|
Event or Problem Description
|
|
It was reported that; patient was put in prone position.This was a regular operation sterilization spread towels, full clear surgical field exposure, conventional cone, open device agent take nail, needle, the establishment of fluoroscopy location is good, choose screw implant, poor position after perspective.Surgeon removed the screw, found the screw detached and never recovered.
|
| |
|