| Model Number |
TV-IL1416100-J |
| Medical Device Problem Codes |
Collapse (1099); Kinked (1339); Leak/Splash (1354); Occlusion Within Device (1423)
|
| Health Effect - Clinical Codes |
Aneurysm (1708); Death (1802); Failure of Implant (1924); Occlusion (1984); Pain (1994); Stenosis (2263); Claudication (2550)
|
| Date of Event |
12/06/2017
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Endologix will continue to investigate the reported event.The patient medical records and imaging studies will be requested for further evaluation by a clinical specialist.A final report will be submitted once the investigation of the reported event has concluded.
|
| |
|
Event or Problem Description
|
|
An ovation ix iliac limb stent graft system was implanted to treat an abdominal aortic aneurysm.Approximately 3 weeks post index procedure, the patient presented with claudication and a cold left foot.A ct was done which showed one limb blocking the flow of the other limb leading to an occlusion.A re-intervention was performed where the physician attempted to balloon the occluded limb, however the patient could not tolerate the procedure and expired.
|
| |
|
Additional Manufacturer Narrative
|
|
Based on the clinical assessment for this event, the most likely cause of the proximal loss of seal was found to be anatomy related, due to the hostile neck anatomy, which likely contributed to the occlusion of the aortic body, right iliac limb stent, and the native vessels down to the popliteal artery.The most likely cause of the occlusion within the stent was found to be the near complete buckling of the right iliac stent at the level of the bifurcation, however, the exact cause of the stent buckling could not be determined.The cause of the death or date of death could not be ascertained due to the lack of medical records surrounding the event.A review of the device quality records shows that the device demonstrated compliance to established procedures and specifications at the time of manufacture.
|
| |
|