| Catalog Number |
400SMTSFT3H08 |
| Medical Device Problem Codes |
Detachment Of Device Component (1104); Retraction Problem (1536); Separation Failure (2547)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
12/15/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Results: the penumbra smart coil (smart coil) embolization coil was detached from its pusher assembly and, therefore, the smart coil was not functionally tested for detachment.Conclusions: evaluation of the returned device revealed the device is returned kinked and folded, and based on the returned condition of the smart coil, the root cause of the reported detachment issue could not be determined.If the device is returned kinked and folded, the penumbra investigator will not be able to determine the reported issue.Evaluation of the returned handle revealed that the handle was functional.The handle was functionally tested on the handle fixture for pull force, throw distance, and passed both specifications without an issue.Penumbra coils are visually inspected during in-process inspection and during quality inspection after manufacturing.Penumbra handles are 100% functionally tested during incoming inspection by quality.
|
| |
|
Event or Problem Description
|
|
The patient was undergoing a coil embolization procedure in the right internal carotid artery-posterior communicating artery (ic-pc) using penumbra smart coils (smart coils) and a penumbra smart coil detachment handle (handle).It was noted that the patient anatomy was slightly tortuous.During the procedure, the physician advanced a smart coil through a non-penumbra microcatheter and attempted to detach it using the handle; however, the coil failed to detach.After four to five failed attempts, the physician bent the smart coil pusher assembly and pulled the pull wire; however, the smart coil did not detach.The physician then decided to retract the smart coil; however, the smart coil unintentionally detached in the microcatheter after being retracted approximately two centimeters back into the microcatheter.Therefore, the physician used the coil''s pusher assembly to push the detached coil into the aneurysm.The procedure was completed using additional coils and the same microcatheter.There was no report of an adverse effect to the patient.
|
| |
|
Search Alerts/Recalls
|
|
|