BOSTON SCIENTIFIC - COSTA RICA (HEREDIA) INTELLATIP MIFI¿ XP TEMPERATURE ABLATION CATHETER; CATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FLUTTER Back to Search Results
Model Number M004PM45000
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).The device was received for analysis.The device was returned in the left curve position.There was dried body fluid on the distal end and the tip had some minor scratches.The seals of all three rings had been compromised.The steering knob and the tension control knob functioned properly on both lock and unlock positions.No abnormal resistance was felt when actuating the steering mechanism.Tip motion was evaluated against curve template.The right curve test passed.The left curve failed to reach the specified template area; the tip did not curve to the left.X-ray showed that a steering wire had retracted to between ring 2 and ring 3.X-rays showed no evidence of guide coil collapse.The distal section was dissected.There was a significant amount of body fluid under r1 and r2 rings and in the interior of the sheath.The kevlar wrap was displaced.One of the steering wires was detached; there was evidence of solder on the center support.The detached steering wire had retracted under the kevlar wrap.The manufacturing batch record review confirmed that the device met all material, assembly and performance specifications.The investigation conclusion is design specification as the device problem was traced back to the design specifications.(b)(4).
 
Event or Problem Description
Reportable based on device analysis completed on 22dec2017.It was reported that the steering mechanism broke.During an ablation procedure to treat atrial fibrillation with an intellatip mifi xp ablation catheter, the physician suspected a steering wire break.The procedure was completed with another of the same device and there were no patient complications.Device analysis revealed that the seals of all three rings had been compromised.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTELLATIP MIFI¿ XP TEMPERATURE ABLATION CATHETER
Common Device Name
CATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FLUTTER
Manufacturer (Section D)
BOSTON SCIENTIFIC - COSTA RICA (HEREDIA)
302 parkway
la aurora
heredia
CS 
MDR Report Key7200255
Report Number2134265-2017-13260
Device Sequence Number13707195
Product Code OAD
UDI-Device Identifier08714729848011
UDI-Public08714729848011
Combination Product (Y/N)N
Initial Reporter StateSC
Initial Reporter CountryUS
PMA/510(K) Number
P020025
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 12/22/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/18/2020
Device Model NumberM004PM45000
Device Catalogue NumberPM4500
Device Lot Number20906446
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/12/2017
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/22/2017
Initial Report FDA Received Date01/17/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-