| Model Number |
72202683 |
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the device was cutting and breaking the sutures.No patient injury was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Visual assessment using a stereo microscope confirmed the movable jaw has a sharp edge on one side.This condition could contribute to fraying and cutting of suture.A review of the complaint records confirmed this failure mode to be isolated in nature.This data indicates that sufficient controls (in-process and final inspections) are in place.No further action is warranted at this time.
|
| |
|
Additional Manufacturer Narrative
|
|
Examination was not possible, as the device was not returned.The investigation could not draw any conclusions about the reported event without the return of the device.
|
| |
|