| Model Number |
610-0001 |
| Medical Device Problem Code |
Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Code |
Loss of Vision (2139)
|
| Date of Event |
12/27/2017
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The explanted corneal inlay was lost during handling and was therefore not available for evaluation.The device history record review of the manufacturing lot for this device was performed and there were no discrepancies or unusual findings related to the reported issue.Inlay shifts in position and decreased vision are listed in the device labeling as known potential risks.(b)(4).
|
| |
|
Event or Problem Description
|
|
The patient underwent implantation of the raindrop corneal inlay in the right eye on (b)(6) 2017.Postoperatively the inlay decentered 2 mm inferiorly.In addition, the patient's best corrected distance visual acuity (bcdva) decreased from 20/20 (preoperatively) to 20/30 immediately prior to inlay explantation on (b)(6) 2017.According to the surgeon, poor fixation during surgery was a contributing factor.At last examination on (b)(6) 2018, bcdva improved to 20/20 with a bandage contact lens in place and the patient was reported as stable.
|
| |
|