• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COVIDIEN MFG DC BOULDER COOL-TIP; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COVIDIEN MFG DC BOULDER COOL-TIP; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES Back to Search Results
Model Number ACT1530
Medical Device Problem Code Temperature Problem (3022)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 12/20/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, intra-operatively, the device¿s heating temperature varies greatly and unable to heat the needle.There was no patient injury and complication associated wit the event.
 
Additional Manufacturer Narrative
Evaluation summary: one device was received for evaluation.This device had been used in the treatment or diagnosis of a patient.A review of the lot number reported indicates that the product was within the assigned expiration date at the time of the reported incident.The returned product did not meet specification as received.Visual inspection found no defects.The reported condition was confirmed.The water circulated normally through the device.The device did not read the temperature properly.The needle was removed.The solder joint at the tip of the thermocouple was found to be broken.The investigation identified the root cause of the reported event to be a poor solder joint.Corrective actions have been implemented to mitigate this issue.If information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COOL-TIP
Common Device Name
ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
COVIDIEN MFG DC BOULDER
5920 longbow dr
boulder CO 80301 3299
MDR Report Key7202920
Report Number1717344-2018-00073
Device Sequence Number164177
Product Code GEI
UDI-Device Identifier10884524000947
UDI-Public10884524000947
Combination Product (Y/N)N
Initial Reporter CountryTW
PMA/510(K) Number
K053290
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional,u
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 10/04/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/07/2020
Device Model NumberACT1530
Device Catalogue NumberACT1530
Device Lot Number70460286X
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/26/2018
Initial Date Received by Manufacturer 12/20/2017
Supplement Date Received by Manufacturer06/12/2018
08/08/2018
Initial Report FDA Received Date01/18/2018
Supplement Report FDA Received Date07/19/2018
10/03/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-