| Model Number |
LSMU1350858 |
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
01/03/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.Receipt of device is expected.The investigation is currently in progress.
|
| |
|
Event or Problem Description
|
|
It was reported that the stent allegedly dislodged from the balloon catheter inside the patient an had to be deployed at the site of dislodgment.Reportedly, another balloon expandable vascular covered stent was used and the procedure was completed.There was no reported patient injury.
|
| |
|
Event or Problem Description
|
|
It was reported that the stent allegedly dislodged from the balloon catheter inside the patient and had to be deployed at the site of dislodgment.There were no anomalies noted during the prep of the device.There was no damage noted to the box, tray or stent cover and the stent cover was easily removed without any issue.The patient was being treated for left iliac stenosis.The lesion was a class c lesion.Access was made via the right and left common femoral artery.The area was calcified and was pre-dilated with multiple balloon catheters prior to inserting a lifestream device and the complaint device.Two terumo 260cm glidewire advantage guidewires and a terumo 7fx25cm destination sheath were used.A lifestream device was inserted however the device was unable to cross the lesion and when the device was attempted to be removed the stent dislodged outside of the sheath.The stent then had to be deployed 10 to 15mm below the intended treatment site.The complaint device was then inserted and there was mild resistance felt when inserting the device into the sheath.The stent dislodged above the intended treatment site when the physician attempted to reposition the stent to pull back 10mm.This left a 10mm gap in between the stents.The stent graft was protected by the sheath when tracking to the lesion.Air evacuation was performed prior to insertion.Reportedly, another balloon expandable vascular covered stent was used and the procedure was completed.Guidewire access was maintained throughout the procedure.There was no reported patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
It was reported that the stent allegedly dislodged from the balloon catheter inside the patient and had to be deployed at the site of dislodgment.There were no anomalies noted during the prep of the device.There was no damage noted to the box, tray or stent cover and the stent cover was easily removed without any issue.The patient was being treated for left iliac stenosis.The lesion was a class c lesion.Access was made via the right and left common femoral artery.The area was calcified and was pre-dilated with multiple balloon catheters prior to inserting a lifestream device and the complaint device.Two terumo 260cm glidewire advantage guidewires and a terumo 7fx25cm destination sheath were used.A lifestream device was inserted however the device was unable to cross the lesion and when the device was attempted to be removed the stent dislodged outside of the sheath.The stent then had to be deployed 10 to 15mm below the intended treatment site.The complaint device was then inserted and there was mild resistance felt when inserting the device into the sheath.The stent dislodged above the intended treatment site when the physician attempted to reposition the stent to pull back 10mm.This left a 10mm gap in between the stents.The stent graft was protected by the sheath when tracking to the lesion.Air evacuation was performed prior to insertion.Reportedly, another balloon expandable vascular covered stent was used and the procedure was completed.Guidewire access was maintained throughout the procedure.There was no reported patient injury.The lot history records have been reviewed with special attention to the manufacturing and inspection of this product.The device was found to have met specifications prior to shipment.No manufacturing anomalies were identified that may have caused or contributed to the reported event.This is the first reported complaint for this lot number and issue to date.There have been no other complaints reported of any type for this lot number.The device was not returned for evaluation.The result of the investigation is inconclusive for the reported failure mode of stent dislodged.The sample was not returned for evaluation.Based upon the available information a definitive root cause cannot be determined.It is also unknown whether patient factors, handling or procedural techniques contributed to the reported event.Based on analysis performed no additional action is required at this time.(b)(4).The ifu states: a device description: 1.Implant the lifestream¿ balloon expandable vascular covered stent is comprised of an electropolished balloon-expandable stent made from (b)(4) stainless steel, encapsulated between two layers of eptfe.B indication for use: the lifestream¿ balloon expandable vascular covered stent is indicated for the treatment of atherosclerotic lesions in common and external iliac arteries.Directions for use: site access and preparation 1.Using standard techniques access the artery and place an introducer sheath or guiding catheter of appropriate inner diameter and a 0.035" (0.89 mm) guidewire across the target lesion.2.Perform diagnostic angiography to confirm site of implantation and measure the reference vessel diameter and lesion length.Covered stent size selection 3.Select a covered stent diameter that is approximately 5%-20% larger than the largest reference vessel diameter at the proximal or distal target site.Refer to the sizing table on the packaging label for appropriate selection of the covered stent diameter and length.Endovascular system preparation 4.Carefully remove the selected device from the package.5.Inspect the covered stent for adherence to the balloon and centered placement in relation to the balloon marker bands.If the covered stent is not centered and/or does not firmly adhere to the balloon, do not use.6.Flush the delivery system guidewire lumen with sterile saline mixture until saline drops from the distal end of the endovascular system.Air evacuation 7.A 20 cc or smaller luer-lock syringe with a minimum of 5 cc¿s sterile saline mixture is recommended for use for aspirating this device.8.With the distal balloon tip pointing down and positioned below the level of the syringe, pull negative pressure until all air is expelled.9.Induce a negative pressure to remove any air from the balloon and inflation lumen.Repeat until all air is expelled.10.Carefully release to neutral.Allow the inflation lumen to fill with the diluted contrast medium and maintain a neutral pressure.Important: do not apply positive pressure to the balloon.11.Attach the prefilled inflation device to the inflation lumen of the catheter hub, ensuring no air bubbles remain at the catheter connection.12.Verify that the covered stent is still centered between the two radiopaque markers on the balloon catheter.Introduction of the endovascular system and placement of the covered stent 13.Advance the endovascular system over the guidewire into the introducer sheath.14.Further advance the endovascular system to the target treatment site within the introducer sheath and position the covered stent across the lesion.Verify that the covered stent is still centered within the balloon marker bands.Slowly retract the introducer sheath / guiding catheter while maintaining the position of the covered stent.Ensure the introducer sheath is retracted far enough to not compromise the balloon expansion and covered stent release.(b)(4).
|
| |
|
Search Alerts/Recalls
|