MAUDE Adverse Event Report: STRYKER ORTHOPAEDICS-MAHWAH V40(TM). FEMORAL HEAD; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ORTHOPAEDICS-MAHWAH V40(TM). FEMORAL HEAD; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED Back to Search Results
Catalog Number 6364-2-036
Medical Device Problem Codes Device Dislodged or Dislocated (2923); Insufficient Device Problem Information (3190)
Health Effect - Clinical Codes Injury (2348); Joint Dislocation (2374)
Date of Event 12/21/2017
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Review of the device history records indicate devices were manufactured and accepted into final stock with no reported discrepancies.There have been no other events for the lot referenced.It was noted that the device is not available for evaluation.Should additional information become available, it will be provided in a supplemental report upon completion of the investigation.
 
Event or Problem Description
It was reported through the submission of an implant sheet that patient's right hip was revised.
 
Additional Manufacturer Narrative
Reported event: an event regarding dislocation involving a metal head was reported.The event was not confirmed.Method & results: device evaluation and results: not performed as the reported device was not returned for evaluation.Clinician review: a medical review was not performed because insufficient information was provided.Device history review: review indicated all devices were manufactured and accepted into final stock with no reported discrepancies.Complaint history review: review indicated there have been no other similar events for the reported lot.Conclusions: it was reported that patient was revised due to dislocation.The event could not be confirmed nor the root cause determined as sufficient information was not provided.Further information such as return of device, operative reports, x-rays, histopathology reports, patient history & follow-up notes are needed to investigate this event further.If additional information and/or device become available, this investigation will be reopened.
 
Event or Problem Description
It was reported through the submission of an implant sheet that patient's right hip was revised.Update- patient was revised due to dislocation.Patient's weight (b)(6) pounds.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
V40(TM). FEMORAL HEAD
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
MDR Report Key7203649
Report Number0002249697-2018-00202
Device Sequence Number1277409
Product Code LPH
UDI-Device Identifier07613153121992
UDI-Public07613153121992
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K121308
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2017
Reporter Type Manufacturer
Report Source health professional,other
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/23/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date09/14/2021
Device Catalogue Number6364-2-036
Device Lot NumberG5816422
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 12/21/2017
Supplement Date Received by Manufacturer05/01/2018
Initial Report FDA Received Date01/18/2018
Supplement Report FDA Received Date05/23/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age63 YR
Patient Weight59
-
-