| Model Number |
TA4535L |
| Medical Device Problem Codes |
Failure to Fire (2610); Mechanics Altered (2984)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
01/02/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter, intra-operatively, the device did not deploy or not come out and stuck inside the cartridge.Opened another reload to complete the procedure.No patient injury has been noted.
|
| |
|
Additional Manufacturer Narrative
|
|
Correction: evaluation summary post market vigilance (pmv) led an evaluation of one device.Visual inspection of the single used loading unit (sulu) noted that all the staples were partially advanced in the cartridge.The staples were in usable unformed condition.A pmv representative instrument was used for all functional testing.The advanced staples in the sulu were reused and reset inside the cartridge.The instrument and sulu were applied to test media with proper staple formation.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.Analysis concluded there were no assembly component related failures.Replication of the advanced staple condition may occur when the firing stroke is not completed during application.The instructions for use of this product state that to fire the instrument, squeeze the handle a second time until it reaches a solid stop and it locks in the back position.If the instrument handle is partially squeezed during firing then released, bleeding may occur as the instrument was not fully fired.The current packaging design for the product does not allow room for advanced staples, thus the condition could not have occurred during assembly or shipment.The root cause of the observed damage was misuse of the product which would have caused or contributed to the reported incident.Should new information become available, the file will be re-opened, and the investigation summary will be amended as appropriate.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Search Alerts/Recalls
|