Device received with intermittent button response due to flattened(escape, up and act ) button dome switch (no crease) and partial unlocked lcd keypad connector noted during visual inspect.Unable to perform all functional testing including the displacement, operating currents, unexpected restart error test, rewind, basic occlusion, occlusion, prime or compromised force sensor system alarm and excessive no delivery test or verify possible under delivery anomaly due to buttons not responding.Device received with cracked battery tube threads, cracked lcd window, minor scratched lcd window, cracked reservoir tube lip, cracked case display window corner, stained end cap sticker, stained address or serial number label, cracked belt clip slot and cracked case reservoir tube window corner.
|