Model Number H7493808012370
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 11/08/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).
 
Event or Problem Description
It was reported that shaft break occurred.The target lesion was located in the left circumflex artery.A 3.75mm x 12mm quantum¿ maverick¿ balloon catheter was advanced to dilate the lesion.However, the shaft broke over 15cm from the hub.The device was completely removed with the guide catheter and the procedure was completed with another with same device.No patient complications were reported and the patient's status was stable.
 
Additional Manufacturer Narrative
Device evaluated by mfr., eval summary attached, method codes, result codes, conclusion codes updated.Device evaluated by manufacturer: returned product consisted of a quantum maverick balloon catheter in three pieces.The balloon was loosely folded.The outer shaft, inner shaft, balloon and tip were microscopically examined.The hypotube shaft was completely separated in two locations.The first hypotube separation is 44.9cm from the hub; the second separation is 32.9cm from the first separation.The fracture faces were oval as if kinked prior to separation.There are numerous hypotube kinks.There was no evidence of any material or manufacturing deficiencies contributing to the damage.The investigation conclusion is operational context as the product meets the design & manufacture specification but due to anatomical/procedural factors encountered during the procedure, performance was limited.(b)(4).
 
Event or Problem Description
It was reported that shaft break occurred.The target lesion was located in the left circumflex artery.A 3.75mm x 12mm quantum¿ maverick¿ balloon catheter was advanced to dilate the lesion.However, the shaft broke over 15cm from the hub.The device was completely removed with the guide catheter and the procedure was completed with another with same device.No patient complications were reported and the patient's status was stable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUANTUM¿ MAVERICK¿
Common Device Name
CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS
Manufacturer (Section D)
BOSTON SCIENTIFIC - MAPLE GROVE
one scimed place
maple grove MN 55311
MDR Report Key7204221
Report Number2134265-2018-00046
Device Sequence Number1485945
Product Code LOX
UDI-Device Identifier08714729392705
UDI-Public08714729392705
Combination Product (Y/N)N
Initial Reporter CountryCN
PMA/510(K) Number
P860019
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,distri
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/28/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/30/2019
Device Model NumberH7493808012370
Device Catalogue Number38080-1237
Device Lot Number0019194466
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/18/2018
Initial Date Received by Manufacturer 12/28/2017
Supplement Date Received by Manufacturer01/31/2018
Initial Report FDA Received Date01/18/2018
Supplement Report FDA Received Date02/26/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-