Model Number N/A
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The 510k status: exempt.(b)(4).This report includes information known at this time.A follow up report will be submitted should additional relevant information become available.
 
Event or Problem Description
The customer reported that the hub off the ultrathane dawson-mueller mac-loc locking loop multipurpose drainage catheter was "leaking once we inserted and because of the leak, the device had to be replaced." there was no report of an adverse event.Additional information has been requested from the customer, but none has yet been provided.The device was received for evaluation; however, as of the date of this report, the investigation is still pending.
 
Additional Manufacturer Narrative
Additional information: device evaluated by mfr.Corrected information: report source.Investigation - evaluation: a review of the complaint history, drawings, dimensional verification, device history record, documentation, instructions for use (ifu), manufacturing instructions, quality control, and functional testing of the returned device was conducted during the investigation.One catheter was returned with the catheter cut several centimeters from the hub.The cap and tubing was unable to be tugged or twisted within the proximal assembly.Biomatter was present.The mac-loc was returned in the locked position.No cracks or defects were observed on the proximal assembly.One thread was present between the mac-loc and cap.Hemostats were used to clamp off the catheter in order to pressurize the proximal assembly.A pressure of 8.5psi was held for approximately 2min with no water leaking.Therefore, the failure mode could not be duplicated.However, it is possible that the hub initially leaked and the biomatter dried causing the leak to stop.A review of production and quality documentation did not observe any specific issues with current manufacturing or quality controls that may have contributed to this incident.Review of the device history record shows no nonconforming events that would contribute to this failure mode.There were no other reported complaints for this lot number.Based on the provided information, inspection of returned product and the results of the investigation, a definitive root cause could not be determined.Measures have been initiated to address this failure mode.Per the quality engineering risk assessment, no further action is required.Appropriate personnel have been notified and monitoring will continue to be performed for similar complaints.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ULTRATHANE DAWSON-MUELLER MAC-LOC LOCKING LOOP MULTIPURPOSE DRAINAGE CATHETER
Common Device Name
GBO CATHETER, NEPHROSTOMY, GENERAL & PLASTIC SURGERY
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
MDR Report Key7204674
Report Number1820334-2018-00053
Device Sequence Number135322
Product Code GBO
UDI-Device Identifier00827002104358
UDI-Public(01)00827002104358(17)200718(10)8074810
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,user f
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 02/20/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue NumberULT12.0-38-25-P-5S-CLDM-HC
Was Device Available for Evaluation? Yes
Initial Date Received by Manufacturer 12/20/2017
Supplement Date Received by Manufacturer02/01/2018
Initial Report FDA Received Date01/18/2018
Supplement Report FDA Received Date02/20/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-