Catalog Number S-65-150-120-P6
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 01/02/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).During processing of this complaint, attempts were made to obtain complete event, patient and device information.The device was not returned for analysis.A review of the lot history record revealed no manufacturing nonconformities that would have contributed to this complaint.Additionally, a review of the complaint history of the reported lot revealed no other incidents.The investigation was unable to determine a conclusive cause for the deployment issue.The elongation likely occurred when the delivery system was being removed with the stent still attached.There is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.
 
Event or Problem Description
It was reported the procedure was performed to treat a lesion in the left superficial femoral artery (sfa) with heavy calcification.The 6.5 x 150 mm supera stent delivery system (sds) was advanced to the lesion.The stent was deployed; however, after deployment, the stent would not come off the delivery system and elongated.The sds was pushed and pulled, locked and unlocked, the handle was twisted and the thumbslide was moved forward and back, but the stent did not release from the sds.The stent was pulled out and a new supera sds was used to complete the procedure.There were no adverse patient effects and no clinically significant delay in the procedure.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SUPERA SELF-EXPANDING STENT SYSTEM
Common Device Name
SELF EXPANDING STENT SYSTEM
Manufacturer (Section D)
AV-TEMECULA-CT
abbott vascular
26531 ynez road
temecula CA 92591 4628
MDR Report Key7207885
Report Number2024168-2018-00488
Device Sequence Number869208
Product Code NIP
UDI-Device Identifier08975370026399
UDI-Public08975370026399
Combination Product (Y/N)N
Initial Reporter StateMO
Initial Reporter CountryUS
PMA/510(K) Number
P120020
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Health Professional
Remedial Action Other
Type of Report Initial
Report Date (Section B) 01/19/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/31/2018
Device Catalogue NumberS-65-150-120-P6
Device Lot Number7010461
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/05/2018
Initial Report FDA Received Date01/19/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age78 YR
Patient Weight66
-
-