| Model Number |
MICL13.2 |
| Medical Device Problem Code |
Device Contamination with Chemical or Other Material (2944)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
No similar complaint type events were reported for units within the same lot.(b)(4).
|
| |
|
Event or Problem Description
|
|
The reporter indicated that the surgeon noticed a microscopic black speck on a micl13.2, -9.5 diopter, implantable collamer lens.The lens was poured out of the vial into the packaging of the cartridge and that's when the speck was noticed.It was reported that the icl had touched nothing but the plunger which was directly opened and used to slide the icl out.The speck was located on the surface of the lens.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: the lens was returned in liquid, in a lens case/vial.Visual inspection found no visible damage to the lens.The lens was inspected under 20x on the microscope and no black spec was noted.Claim # (b)(4).
|
| |
|