|
Additional narrative: patient weight was not provided for reporting.Date of event is unknown.(b)(4).The device was received and the product evaluation is in progress.No conclusion can be drawn.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Device history records review was completed for part# 206.035, lot# l240034.Manufacturing location: (b)(4), release to warehouse date: dec 13, 2016.Review of the device history records showed that there were no issues during the manufacture of the product that would contribute to this complaint condition.Product development investigation was completed.A device history record (dhr) review, device inspection, and x-ray review were performed as part of this investigation.The returned cancellous bone screw ø 4.0mm, l 35mm was visually checked and no damage could be detected.The review of the device history record revealed that this device was manufactured in december 2016 according to the specifications.The part conformed to dimensional specifications at the time of manufacturing and passed inspection requirements with no non-conformity specific to the part reported.The lot went through the required steps during the inspection at the time of manufacturing and the dhr records showed no issues concerning the material or material conditioning.The visual inspection of the cancellous bone screw did not show any abnormalities neither any defects could be found.Provided x-rays were reviewed.No issues with the screws were identified.Thus, the complaint condition for this device could not be confirmed or replicated.There is no indication that a design or manufacturing issue contributed to the complaint.No malfunctions were observed during the course of this investigation.As no malfunction was identified, no further investigation was performed.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|