| Model Number |
9004 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Reocclusion (1985)
|
| Date of Event |
12/06/2016
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Analysis: the sample was not returned from the user facility; therefore, a device evaluation was unable to be performed.A lot history review revealed 5 additional reocclusion complaints were associated with this lot.A review of the device history record (dhr) indicates the lot was manufactured to specification.Conclusion: the actual sample was not received for evaluation.The dhr found nothing to indicate a manufacturing related cause for this event.The investigator assessed the event was possibly related to the study device and procedure.Based on the instructions for use (ifu), the occurrence of reocclusion is an inherent risk of any pta procedure, and has been reported in clinical trials of drug coated balloons.If additional information becomes known to the manufacturer, a supplemental report will be provided with all relevant information.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
|
| |
|
Event or Problem Description
|
|
It was reported through a clinical registry that during the index procedure, a lutonix percutaneous transluminal drug coated balloon (dcb) dilatation catheter was used to treat the target lesion located in the left mid superficial femoral artery (sfa).Approximately 12 months after the index procedure, the patient¿s left mid sfa vessel was reportedly reoccluded.The patient was placed on a conservative medication regiment and no additional interventions have been performed at this time.The investigator assessed that the events were possibly related to the study device and procedure.The sample was discarded by the user facility and is not available for further evaluation.No adverse patient effects were reported.
|
| |
|
Event or Problem Description
|
|
It was reported through a clinical registry that during the index procedure, a lutonix percutaneous transluminal drug coated balloon (dcb) dilatation catheter was used to treat the target lesion located in the left mid superficial femoral artery (sfa).Approximately 12 months after the index procedure, the patient¿s left mid sfa vessel was reportedly reoccluded.The patient was placed on a conservative medication regiment and no additional interventions have been performed at this time.The investigator reassessed as not related to index procedure , and the study device; therefore, the event has been deleted.The sample was discarded by the user facility and is not available for further evaluation.No adverse patient effects were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The purpose of this supplement is to inform fda that the event was reassessed as not related to index procedure/study device and deleted.H3 analysis: the sample was not returned from the user facility; therefore, a device evaluation was unable to be performed.A lot history review revealed 5 additional reocclusion complaints were associated with this lot.A review of the device history record (dhr) indicates the lot was manufactured to specification.Conclusion: the actual sample was not received for evaluation.It is known that the patient¿s left mid sfa vessel was reportedly reoccluded.The patient was placed on a conservative medication regiment and no additional interventions have been performed at this time.The investigator assessed that the event was not related to the study device or procedure.No adverse patient effects were reported.H11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|