• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CONMED Y-KNOT RC ALL-SUTURE ANCHOR, WITH TWO NO. 2 ( 5 METRIC) HI-FI SUTURES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CONMED Y-KNOT RC ALL-SUTURE ANCHOR, WITH TWO NO. 2 ( 5 METRIC) HI-FI SUTURES Back to Search Results
Catalog Number YRC02
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 10/10/2016
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The reported driver was not returned to conmed for evaluation; therefore, this alleged product failure is unable to be verified.The most probable cause for this breakage is excessive force applied on the device during use.A review of the manufacturing documents has verified the devices were produced per current and approved procedures and material specifications.Non-conformances regarding the product's identity, quality, safety, effectiveness or performance were not identified during manufacture that would contribute to this issue.A two-year historical complaint review revealed one similar complaint for this device family.In that same timeframe, (b)(4) devices have been manufactured and shipped worldwide, making the rate of occurrence of this failure (b)(4) percent.A risk analysis was performed and found acceptable.The instructions for use advise the user of the following.Inspect all instruments prior to use to ensure that are in good physical condition and function properly.There should be no loose, broken or misaligned parts.Exercise care in the use of these devices to minimize side or bending loads.Do not use excessive force on instruments to avoid damage or breakage during use.Avoid unintended contact with other surgical instruments during use to prevent damage or breakage.This incident type will continue to be monitored within the complaint system to assure patient safety.
 
Event or Problem Description
This incident was reported to conmed on (b)(4) 2016 and deemed nonreportable on 31-oct-2016.On 08-jan-2018, a historical complaint review was pulled for this product code and this incident was reassessed.A customer reported during an arthroscopic shoulder surgery, an anchor broke and a tip was left in the pilot hole.The surgeons could not retrieve this "tip" as the patient's bone was hard.The procedure was completed using an alternative device.No patient injury was reported.This incident has been deemed reportable on the basis of a device malfunction where a non-implantable device component remained in the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Y-KNOT RC ALL-SUTURE ANCHOR, WITH TWO NO. 2 ( 5 METRIC) HI-FI SUTURES
Common Device Name
Y-KNOT RC ALL-SUTURE ANCHOR
Manufacturer (Section D)
CONMED
11311 concept blvd
largo FL 33773
MDR Report Key7210530
Report Number1017294-2018-00006
Device Sequence Number16390629
Product Code MBI
UDI-Device Identifier10845854080542
UDI-Public(01)10845854080542(17)210510(30)1(10)748403
Combination Product (Y/N)N
Initial Reporter CountrySW
PMA/510(K) Number
K133224
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,other,user facility
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 01/22/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/10/2021
Device Catalogue NumberYRC02
Device Lot Number748403
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/08/2018
Initial Report FDA Received Date01/22/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age52 YR
-
-