Model Number MMT-751NAP
Health Effect - Clinical Code Hyperglycemia (1905)
Date of Event 01/14/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
 
Event or Problem Description
Customer reported via phone call for high blood glucose.Customer¿s blood glucose was 453 mg/dl.Customer stated that she has a low battery alarm and has been trying to get the battery cap unscrewed and cant.Customer stated that she has been trying for at least 8 hours and can¿t get anything open.Customer declined high blood glucose troubleshooting as she knows why her high happened.The insulin pump will be returned for analysis.
 
Additional Manufacturer Narrative
Device passed the operating currents, self test and displacement test.No low battery alert noted during test.Device received with cracked battery tube threads, stripped battery cap coin slot, cracked lcd window, minor scratched lcd window, cracked reservoir tube lip, cracked case display window corner, stained end cap sticker and stained address or serial number label.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
530G INSULIN PUMP MMT-751NAP
Common Device Name
ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
MDR Report Key7210797
Report Number3004209178-2018-45200
Device Sequence Number16334578
Product Code OZO
UDI-Device Identifier00643169736092
UDI-Public(01)00643169736092
Combination Product (Y/N)N
Initial Reporter StateKY
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Patient
Type of Report Initial,Followup
Report Date (Section B) 05/30/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Model NumberMMT-751NAP
Device Catalogue NumberMMT-751NAP
Device Lot NumberA6751NAPJ
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/01/2018
Initial Date Received by Manufacturer 01/14/2018
Supplement Date Received by Manufacturer05/11/2018
Initial Report FDA Received Date01/22/2018
Supplement Report FDA Received Date05/30/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age31 YR
Patient Weight130
-
-