| Model Number |
MT22949 |
| Medical Device Problem Code |
Inappropriate or Unexpected Reset (2959)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
01/01/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).
|
| |
|
Event or Problem Description
|
|
Dexcom was made aware on (b)(4) 2018 that on (b)(6) 2018 the receiver initialized without a manual restart.No additional event or patient information is available.No product or data were provided for evaluation.The confirmation of the complaint is undetermined.A root cause could not be determined.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).
|
| |
|
Event or Problem Description
|
|
Dexcom was made aware on (b)(6) 2018 that on (b)(6) 2018 the receiver initialized without a manual restart.No additional event or patient information is available.No data was provided for evaluation.The receiver was returned for evaluation.External visual inspection was performed and passed.A receiver charge and boot was performed and failed due to perpetual rebooting.The receiver case was opened, and the ui pcba board was detached from the receiver.A log review was unable to be performed.A functional test could not be performed.Confirmation of the complaint was undetermined.The probable cause could not be determined.
|
| |
|
Search Alerts/Recalls
|