Model Number H7493911416250
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Chest Pain (1776); Dyspnea (1816); Cardiac Enzyme Elevation (1838); Myocardial Infarction (1969); Reocclusion (1985); Tachycardia (2095)
Date of Event 07/12/2017
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Device evaluated by mfr: it is indicated that the device will not be returned for evaluation.If there is any further relevant information obtained, a supplemental medwatch will be filed.(b)(4).
 
Event or Problem Description
(b)(6) clinical study.It was reported that myocardial infarction and restenosis occurred.In (b)(6) 2012, the subject presented with unstable angina (braunwald classification: iib) and was referred for cardiac catheterization.Target lesion #1 was a de novo lesion located in mid right coronary artery (rca) extended to distal rca with 75% stenosis and was 58 mm long with a reference vessel diameter of 3.0 mm.Target lesion #1 was treated with pre-dilatation and placement of 2.75 mm x 38 mm and 2.75 mm x 24 mm promus element plus stents in an overlapping fashion.Following post dilatation residual stenosis was 25%.Target lesion #2 was a de novo lesion located in the right posterior descending artery (r-pda) with 75% stenosis and was 12 mm long with a reference vessel diameter of 2.5 mm.Target lesion #2 was treated with direct placement of a 2.50 mm x 16 mm promus element plus stent.Following post dilatation residual stenosis was 25%.The subject was discharged on aspirin and clopidogrel.In (b)(6) 2017, the subject presented with the complaints of chest pain associated with shortness of breath and was hospitalized on the same day.Chest x ray revealed interstitial thickening with no acute infiltrate.No vascular congestion or pneumothorax was seen.There was a mild cardiomegaly noted.The subject¿s troponin level was noted to be elevated.Electrocardiography (ekg) revealed sinus tachycardia with premature atrial complexes with aberrant conduction, left bundle branch block.The subject was referred for cardiac catheterization.The 90% stenosis located in r-pda was treated with plain balloon angioplasty following which 20% residual stenosis and timi 3 flow were noted.The event was considered as resolved.
 
Additional Manufacturer Narrative
(b)(4).
 
Event or Problem Description
It was further reported that the study stent was patent and the reported myocardial infarction was considered not related to the device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROMUS ELEMENT¿ PLUS
Common Device Name
STENT, CORONARY, DRUG-ELUTING
Manufacturer (Section D)
BOSTON SCIENTIFIC - GALWAY
MDR Report Key7211538
Report Number2134265-2018-00264
Device Sequence Number13742759
Product Code NIQ
UDI-Device Identifier08714729807834
UDI-Public08714729807834
Combination Product (Y/N)N
Initial Reporter StateGA
Initial Reporter CountryUS
PMA/510(K) Number
P110010
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2012
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/29/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date08/27/2013
Device Model NumberH7493911416250
Device Catalogue Number39114-1625
Device Lot Number0015584585
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 12/29/2017
Supplement Date Received by Manufacturer01/25/2018
Initial Report FDA Received Date01/22/2018
Supplement Report FDA Received Date02/15/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age72 YR
-
-