• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC - GALWAY GLADIATOR¿; CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC - GALWAY GLADIATOR¿; CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL Back to Search Results
Model Number H74939207080870
Device Problem Material Rupture (1546)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 12/29/2017
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Device evaluated by mfr.: the complaint device was returned for analysis.The balloon was unfolded which indicated it had been subjected to positive pressure.No issues were identified with the balloon material that could have contributed to the complaint incident.A visual and microscopic examination found no issues with the markerbands or tip of the device that could have contributed to the complaint incident.A visual and microscopic examination identified a pinhole leak in the shaft of the device located approximately 90mnm proximal of the proximal balloon bond.No other issues were noted with the device during analysis.The investigation conclusion is operational context as the product meets the design & manufacture specification but due to anatomical/procedural factors encountered during the procedure, performance was limited.(b)(4).
 
Event Description
It was reported that catheter rupture occurred.A 8.0 x80 75cm gladiator¿ balloon catheter was used to dilate the lesion.However, the catheter ruptured.The procedure was completed with another of the same device.No patient complications were reported and the patient's status was fine.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GLADIATOR¿
Type of Device
CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL
Manufacturer (Section D)
BOSTON SCIENTIFIC - GALWAY
Manufacturer (Section G)
BOSTON SCIENTIFIC - GALWAY
Manufacturer Contact
sonali arangil
one scimed place
maple grove, MN 55311
7634941700
MDR Report Key7211549
MDR Text Key97978721
Report Number2134265-2018-00091
Device Sequence Number1
Product Code LIT
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K113681
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 12/29/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/22/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date10/08/2020
Device Model NumberH74939207080870
Device Catalogue Number3920708087
Device Lot Number0021232755
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/04/2018
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/29/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/09/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-