| Model Number |
X SERIES |
| Medical Device Problem Code |
Failure of Device to Self-Test (2937)
|
| Health Effect - Clinical Code |
No Information (3190)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was returned to zoll medical corporation for evaluation.During disassembly of the device to determine root cause, the technician observed damage consistant with mis-handling and user tampering.It cannot be determined if our observations are pre or post event.The device's defib cable receptacle and monitor board were replaced to remedy the damage.The device was recertified and returned to the customer.Analysis of reports of this type has not identified an increase in trend.
|
| |
|
Event or Problem Description
|
|
Complainant alleged that while attempting to defibrillate a patient (age & gender unknown), the device failed to discharge.Complainant did not indicate that there was any adverse effect to the patient due to the reported malfunction.
|
| |
|
Search Alerts/Recalls
|