• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. 530G INSULIN PUMP MMT-551RNAS; ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. 530G INSULIN PUMP MMT-551RNAS; ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND Back to Search Results
Model Number MMT-551RNAS
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 09/11/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The insulin pump was received with detached end cap, cracked reservoir tube lip, cracked battery tube threads, cracked belt clip slot and minor scratched lcd window.The insulin pump was unable to perform the basic occlusion test, occlusion test, prime test and excessive no delivery test due to detached end cap.
 
Event or Problem Description
The customer reported via phone call that the bottom of the insulin pump was detached and would not stick back in.The customer's blood glucose was 206 mg/dl at the time of incident.The insulin pump was returned for analysis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
530G INSULIN PUMP MMT-551RNAS
Common Device Name
ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
MDR Report Key7212869
Report Number3004209178-2018-45405
Device Sequence Number1213367
Product Code OZO
UDI-Device Identifier00643169507951
UDI-Public(01)00643169507951
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Patient
Type of Report Initial
Report Date (Section B) 01/23/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberMMT-551RNAS
Device Catalogue NumberMMT-551RNAS
Device Lot NumberA3551RNASJ
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/19/2017
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/12/2017
Initial Report FDA Received Date01/23/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age73 YR
Patient Weight177
-
-