• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT MEDICAL OPTICS TECNIS 1 MULTIFOCAL; MULTIFOCAL IOLS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT MEDICAL OPTICS TECNIS 1 MULTIFOCAL; MULTIFOCAL IOLS Back to Search Results
Model Number ZLB00
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Halo (2227); No Code Available (3191)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Date of event: unknown, not provided, but the best estimate date is between (b)(6) 2017 and (b)(6) 2018.(b)(4).All pertinent information available to abbott medical optics has been submitted.
 
Event or Problem Description
It was reported that a zlb00 10.0 diopter intraocular lens (iol) was implanted in the patient's left eye (os) on (b)(6) 2017.It was later explanted on (b)(6) 2018, because the patient complained of halos and was overall dissatisfied with the lens that was implanted.The symptoms also significantly interfered with regular activities.The lens was replaced with the same diopter, but different model z9002.There was no incision enlargement, vitrectomy, or sutures used.Also, there was no post-operative patient injury.No additional information was provided.
 
Additional Manufacturer Narrative
Device available for evaluation? yes.Returned to manufacturer on: 2/16/2018.Device returned to manufacturer? yes.Device evaluation: the intraocular lens (iol) was returned to the manufacturer.Visual inspection with the unaided eye shows the product is cut in pieces, most probably to make explant possible.Considering the condition of the lens additional analysis is not possible.The complaint cannot be verified.Manufacturing record review: manufacturing records were reviewed and the device was manufactured according to specification.A search on complaints history revealed no similar complaints were received for this production order number to date.Labeling review: the directions for use (dfu) was reviewed.The dfu adequately provides instructions and precautions for the proper use and handling of the product.Conclusion: based on the investigation results there is no indication of a product quality deficiency.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS 1 MULTIFOCAL
Common Device Name
MULTIFOCAL IOLS
Manufacturer (Section D)
ABBOTT MEDICAL OPTICS
santa ana CA 92705
MDR Report Key7213419
Report Number9614546-2018-00083
Device Sequence Number869093
Product Code MFK
UDI-Device Identifier05050474560475
UDI-Public(01)05050474560475(17)190318
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2018
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/07/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Physician
Device Expiration Date03/18/2019
Device Model NumberZLB00
Device Catalogue NumberZLB00U0100
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/16/2018
Initial Date Received by Manufacturer 12/28/2017
Supplement Date Received by Manufacturer04/10/2018
Initial Report FDA Received Date01/23/2018
Supplement Report FDA Received Date05/07/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-