|
(b)(4).During processing of this complaint, attempts were made to obtain complete event, patient and device information.(b)(4).Evaluation summary: visual and functional inspection was performed on the returned device.The reported crack was not confirmed; however, the inflation issue was confirmed due to a leak in the hub.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history of the reported lot revealed no other incidents.Av conducted root cause analysis and determined the issue may be related to a potential product quality issue.The issue is being addressed per internal operating procedures.Av will continue to trend the performance of these devices.
|