Model Number S7
Device Problems Unintended Arm Motion (1033); Material Integrity Problem (2978); Unintended Movement (3026)
Patient Problem No Patient Involvement (2645)
Event Date 12/29/2017
Event Type  malfunction  
Manufacturer Narrative
No patient information provided as no patient was involved in this concern.A medtronic representative went to site to test the equipment.Representative was able to replicate the reported issue.Representative reported that a slipping issue was observed at the mayfield attachment end ball and socket joint.Usage wear was also observed on the vertek arm.
 
Event Description
A medtronic representative on behalf of site reported that while outside or procedure, the vertek arm was was loose and could not remain in rigid position.There was no patient present at the time of the issue.
 
Manufacturer Narrative
Product, unique device identification (udi) and associated fields updated to proper value.
 
Manufacturer Narrative
The fda codes were updated to reflect current coding guidelines.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STEALTHSTATION S7 SYSTEM
Type of Device
NEUROLOGICAL STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer Contact
peter verhey
attn:product quality experienc
826 coal creek circle
louisville, CO 80027-9710
7208902515
MDR Report Key7214163
MDR Text Key98047110
Report Number1723170-2018-00349
Device Sequence Number1
Product Code HAW
UDI-Device Identifier00613994450944
UDI-Public00613994450944
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K050438
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Medical Equipment Company Technician/Representative
Type of Report Initial,Followup,Followup
Report Date 05/30/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/23/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Physician
Device Model NumberS7
Device Catalogue Number9733927
Is the Reporter a Health Professional? No
Date Manufacturer Received01/24/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/28/2010
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-