• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: HEARTWARE, INC. HEARTWARE VENTRICULAR ASSIST SYSTEM - CONTROLLER DC ADAPTER; VENTRICULAR (ASSISST) BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

HEARTWARE, INC. HEARTWARE VENTRICULAR ASSIST SYSTEM - CONTROLLER DC ADAPTER; VENTRICULAR (ASSISST) BYPASS Back to Search Results
Model Number 1440
Medical Device Problem Code Power Problem (3010)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 01/09/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This device is used for treatment not diagnosis.The ventricular assist system is indicated for use as a bridge to cardiac transplantation and destination therapy in patients who are at risk of death from refractory end-stage left ventricular heart failure.The system is designed for in-hospital and out-of-hospital settings.
 
Event or Problem Description
It was reported that the controller dc adapter was not providing power.The adapter was exchanged.No patient complications have been reported as a result of this event.
 
Additional Manufacturer Narrative
Product event summary: the controller dc adapter was returned for evaluation.Various analyses were conducted and reviewed in order to evaluate the performance of the device in relation to the reported event.Failure analysis of the returned device revealed that the controller dc adapter passed visual examination and functional testing.The controller dc adapter was able to adequately provide power to a test controller.As a result, the reported event could not be confirmed or duplicated during testing.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEARTWARE VENTRICULAR ASSIST SYSTEM - CONTROLLER DC ADAPTER
Common Device Name
VENTRICULAR (ASSISST) BYPASS
Manufacturer (Section D)
HEARTWARE, INC.
14400 nw 60th ave
miami lakes FL 33014
MDR Report Key7214212
Report Number3007042319-2018-00364
Device Sequence Number988114
Product Code DSQ
UDI-Device Identifier00888707000260
UDI-Public00888707000260
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
P100047
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date (Section B) 05/02/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model Number1440
Device Catalogue Number1440
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/12/2018
Initial Date Received by Manufacturer 01/09/2018
Supplement Date Received by Manufacturer04/11/2018
Initial Report FDA Received Date01/23/2018
Supplement Report FDA Received Date05/02/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age54 YR
-
-