| Catalog Number |
9791221 |
| Medical Device Problem Code |
Migration or Expulsion of Device (1395)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The device was not returned to the manufacturer for evaluation.Therefore, we are unable to determine the definitive cause of the reported event.X-ray review post-op x-rays for three level anterior cervical discectomy and fusion was provided.We are unable to determine levels from the given images.There is a screw backed out from the inferior level of the construct.There is a paucity of bone present in the grafts.This may be caused by failure of fusion or failure to properly lock plate over screws.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported that patient underwent a surgery.Post-op, a screw backed out.As a result, the patient underwent a revision surgery, in which the backed out screw was replaced with a rescue screw.No patient complications were reported as a result of the event.
|
| |
|