| Model Number |
710200S |
| Medical Device Problem Code |
Defective Alarm (1014)
|
| Health Effect - Clinical Codes |
Low Blood Pressure/ Hypotension (1914); Blood Loss (2597)
|
| Date of Event |
12/23/2017
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Exemption number e2015012.B.Braun medical inc (importer) is submitting this report on behalf of b.Braun avitum ag (manufacturer).This report has been identified as b.Braun avitum internal report # (b)(4).A b.Braun us technician inspected the machine at the user facilty.The treatment trend report for the incident date of (b)(6) 2017 were reviewed.The report indicates the machine did alarm for venous and arterial pressures.The audible alarms were verified to be working when triggered.The investigation is on-going at this time.A follow-up report will be submitted when the investigation results become available.
|
| |
|
Event or Problem Description
|
|
As reported by the user facility: reported that during a therapy the venous access became dislodged from the patient access and the machine did not alarm.The nurse called paramedics due to blood loss and hypotension.
|
| |
|
Event or Problem Description
|
|
As reported by the user facility: reported that during a therapy the venous access became dislodged from the patient access and the machine did not alarm.The nurse called paramedics due to blood loss and hypotension.Additional information was received on 2/15/2018.The venous needle dislodged two (2) hours and thirty (30) minutes after the start of therapy.The blood loss was approximately 300-400ml.It was indicated there was no damage at the connector of the vascular access or the bloodline system.
|
| |
|
Additional Manufacturer Narrative
|
|
Exemption number e2015012.(b)(4).A b.Braun us technician inspected the machine at the user facility.The treatment trend report for the incident date of (b)(6) 2017 were reviewed.The report indicates the machine did alarm for venous and arterial pressures.The audible alarms were verified to be working when triggered.The dialog+ dialysis machine involved in this event was inspected by the b.Braun technician.He reviewed the trend file onsite during the service call and determined that both the venous and arterial alarms were triggered.It was confirmed that there was no malfunction.The dialog+ machine operated as intended.A dialysis machine can detect a venous needle dislodgement if the venous pressure drops below the lower venous pressure limit, as in the current case.If only a small or no pressure change may occur and the venous pressure does not fall below the lower limit, the venous pressure monitor might not detect and alarm a venous needle dislodgement.Therefore, the operator has to monitor the access side carefully.In addition, the lower alarm limit for venous pressure monitoring should be set as closely to the current value as possible (e.G.20 mm hg).These characteristics of a dialysis machine and the consequences are outlined in the guideline for safe operation of medical equipment used for hemodialysis treatments (iec/tr 62653 edition 1.0 2012-06) chapter 5.3.1 and 5.3.3.Therefore, the instructions for use of the dialog+ dialysis machine warns the user in chapter 5.3 as follows: - risk to patient due to blood loss if cannulas get disconnected or slip out! - standard monitoring function of the dialysis machine cannot safely detect that such a situation has arisen! - ensure that the access to the patient always remains fully visible during therapy.- ensure that cannulas are adequately fixed.- regularly check patient access.- venous lower limit should preferably be > 0 mmhg.Since there was no malfunction of the dialog+ dialysis machine complained in the current case, no further measures will be initiated.If additional pertinent information becomes available a follow-up report will be filed.
|
| |
|