Consumer reported complaint for high blood glucose test results.Wife is calling on behalf of the customer.The customer is concerned with test results from results obtained of 254 and 206 mg/dl.The customer's expected fasting blood glucose test result range is 120 - 140 mg/dl.The customer feels well and did not report any symptoms.Medical attention is not reported as a result of the actual blood glucose results.During the call on (b)(6) 2018, a back to back blood test was performed by the customer non-fasting and produced test results of 187 mg/dl and 200 mg/dl using truetrack meter.The product is stored according to specification in the dining room.The test strip lot manufacturer's expiration date is 05/17/2020 and open vial date is (b)(6) 2017.The meter memory was reviewed for previous test result history (date/time not set): result 1 :161mg/dl date (b)(6) 2018 time:2:24 am fasting, result 2 :163mg/dl date: (b)(6) 2018 time:3:00 am fasting, result 3 :206mg/dl date: (b)(6) 2018 time:1:41 am fasting, result 4 :254mg/dl date: (b)(6) 2018 time:7:20 am fasting, result 5 :191mg/dl date: (b)(6) 2018 time:2:31 am fasting.The customer is calling because he feels that the results he is getting from the meter are reading high.The customer stated that the date/time has not been setup (wrong date/time).
|